ARTIS Q, ARTIS Q. ZEN, and ARTIS PHENO (2018-06-22) Jun 22, 2018 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number 105012 Manufacturer Siemens Healthcare GmbH Henkestr. 127 Erlangen 91052 GERMANY B. ARTIS Q. ZEN 10848353 164101 Manufacturer Siemens Healthcare GmbH Henkestr. 127 Erlangen 91052 GERMANY C. ARTIS PHENO 10849000 More than 10 numbers, contact manufacturer. Manufacturer Siemens Healthcare GmbH Henkestr. 127 Erlangen 91052 GERMANY A. ARTIS Q 10848280
Affected products
A. ARTIS Q
B. ARTIS Q. ZEN
C. ARTIS PHENO
In case of a short-circuit within the x-ray tube a leakage current might cause an overload of the x-ray generator. If this current leakage is undetected it may damage the generator by overheating. X-ray will not be possible.