Health Products

Lysercell WDF (2018-06-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Y7034
Manufacturer
Sysmex Reagents America
Two Sysmex Way
Mundelein
60060
Illinois
UNITED STATES
Lysercell WDFZA900001

Introduction

Affected products

Lysercell WDF

Reason

During an audit of manufacturing records for LYSERCELL WDF, it was discovered that a nonconforming raw material was incorrectly used to manufacture lot Y7034. LYSERCELL WDF is a lysing reagent that is used together with Fluorocell WDF stain to provide the counts and percentages of Neutrophils (Neut % and Neut#), Lymphocytes (lymph% and lymph#), Monocytes (Mono% and Mono#), and Eosinophils (Eo% and Eo#) on the Sysmex XN-series and XN-L Series of automated hematology analyzers. LYSERCELL WDF hemolyzes red blood cells and the Fluorocell WDF dye stains the white blood cells.

If the LYSERCELL WDF lot Y7035 is used, there may be an increase in false positive differential flags on patient samples. The specific flags are interpretive messages "Blasts/Abn Lympho" and "Atypical Lymph", which prompt the XN-Series operator to verify differential results by performing manual smear review. The quality control and reportable absolute cell counts and percentages are not affected.