Optima CT540 Series CT Scanner System (2018-04-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
00000332406HM8
Manufacturer
GE Hangwei Medical Systems Co. Ltd.
WEST AREA OF BLDG. NO.3, NO.1 YONGCHANG NORTH ROAD, BEIJING, ECONOMIC & TECHNOLOGICAL DPVMT ZONE
Beijing
100176
CHINA
Optima CT540 Series CT Scanner System5432539

Introduction

Affected products

Optima CT540 Series CT Scanner System

Reason

The following issue is being addressed by GE healthcare file no. FMI 26859: GE healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC (volts alternating current) to service engineers working inside the gantry. This issue does not affect patients or operators external to the equipment.