Health Products

Platinium SonR CRT-D (2018-07-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
641DL093
Manufacturer
Sorin Group Italia S.R.L.
Via Crescentino, SN,
Saluggia, Vercelli,
13040
ITALY
Platinium SonR CRT-D1811

Introduction

Affected products

Platinium SonR CRT-D

Reason

On a subset of Platinum ICD and CRT-D devices, a specific hardware configuration was identified as potentially defective over time, leading to overconsumption, immediately followed by loss of pacing and sensing capabilities in all cavities. As a result of the loss of sensing capability, the device cannot identify an arrhythmia that would require a defibrillation shock therapy.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime