Health Products

ThromboType (2018-07-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
3005154
Manufacturer
Immucor GTI Diagnostics, Inc.,
20925 Crossroads Circle,
Waukesha
53186
UNITED STATES
ThromboTypeTHROMBOTYPE

Introduction

Affected products

ThromboType

Reason

As a result of a stability failure at the 15month test point, ThromboType kit, Lot 3005154, has the potential to generate an incorrect result or no result with HPA-1 and/or HPA-3. The HPA-1a and HPA-3a allele specific primers for the affected ThromboType lot are experiencing low to no reactivity. OOS 18-98 was initiated to investigate weak and/or absent bands in HPA-1a and HPA-3a during the initial testing on 05/30/18. The failure mode observed in OOS 18-98 was confirmed during repeat testing on 06/11/18.