The affected HistoCore SPECTRA ST Stainers have an incorrect specification against safety standard ISO61010-1. Should a single fault error occur, a mains wire can contact a metal plate which is accessible to the user. This could lead to mains power on the metal plate resulting in hazardous voltage to the user.
Though the probability of this fault is low, safety standard ISO61010-1 does require a 2 layer safety level. Corrective actions are required to prevent this error from occurring and to ensure the 2 layer safety level is established.