Health Products

HistoCore SPECTRA ST Stainer (2018-05-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
148
Manufacturer
LEICA BIOSYSTEMS NUSSLOCH GMBH
HEIDELBERGER STREET 17-19
NUSSLOCH
69226
GERMANY
HistoCore SPECTRA ST StainerNot applicable

Introduction

Affected products

HistoCore SPECTRA ST Stainer

Reason

The affected HistoCore SPECTRA ST Stainers have an incorrect specification against safety standard ISO61010-1. Should a single fault error occur, a mains wire can contact a metal plate which is accessible to the user. This could lead to mains power on the metal plate resulting in hazardous voltage to the user.

Though the probability of this fault is low, safety standard ISO61010-1 does require a 2 layer safety level. Corrective actions are required to prevent this error from occurring and to ensure the 2 layer safety level is established.