Health Products

Cobas p 612 pre-analytical system LCP1 (2018-07-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Software version 2.0 and higher
Manufacturer
Roche Diagnostics GMBH
Sandhoferstrasse 116
Mannheim
68305
GERMANY
Cobas p 612 pre-analytical system LCP107563116001

Introduction

Affected products

Cobas p 612 pre-analytical system LCP1

Reason

Sample material may potentially come in contact with the pipetting nozzle during operation when physically configured to use non-filter tips causing contamination which may result in potential false positive results depending on the sensitivity of the analytical technology.