Alinity ci-series System Control Module (SCM) (2018-07-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
S/N: SCM01196
Manufacturer
Abbott GmbH and Co. K.G.
Max Planck Ring 2
Wiesbaden
65205
GERMANY
Alinity ci-series System Control Module (SCM)03R70-01

Introduction

Affected products

Alinity ci-series System Control Module (SCM)

Reason

Abbott has identified the following issues with the Alinity CI-series which may present a potential performance issue in Alinity CI-series Software version 2.10. Abbott is releasing Alinity CI-series Software version 2.50 to correct these issues.
1. Removing racks during the Maintenance state is not detected by the system. Issues with orders may occur when reloading a rack with samples and returning to Running state.
2. Orders created using more than one of the I-series, C-series, or calculated options to select assays will only include the assays selected in the last filtered view, if the order is saved prior to returning to the All option.
3. Enabling an assay when the processing module is in the Running or Processing status causes the assay to be re-installed.
4. A sample with 270 test orders remained in the Scheduled status and would not process.
5. Some maintenance and diagnostic procedures can appear to have completed successfully when there was insufficient inventory.