Health Products

Reliance 1227 Cart and Utensil Washer/Disinfector (2018-05-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
S/N: 3605113001Reliance 1227 Cart and Utensil Washer/Disinfector
Manufacturer
Steris Canada Corporation
490 Boul. Armand-Paris
Quebec
G1C 8A3
Quebec
CANADA
FW03-101

Introduction

Affected products

Reliance 1227 Cart and Utensil Washer/Disinfector

Reason

STERIS has become aware that the Reliance 1227 Cart and Utensil Washer/Disinfector's "Chemical Low Level" alarm, intended to prevent the user from initiating a cycle when a low chemical level situation occurs, does not operate as intended. Currently, if a low chemical situation occurs, the alarm will only be generated at the unit's next power-up or when accessing service mode. Assuming pre-cleaning and post-processing visual inspection activities are completed in accordance with Good Hospital Practice, the Reliance 1227 Cart Washer's thermal disinfecting properties are not affected by the software issue identified.