Stat Profile Prime CCS Comp Analyzer (2017-06-09)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
P07115190
Manufacturer
Nova Biomedical Corporation
200 Prospect St
Waltham
02454
Massachusetts
UNITED STATES
Stat Profile Prime CCS Comp Analyzer52857

Introduction

Affected products

Stat Profile Prime CCS Comp Analyzer

Reason

When a patient result is in a pending state, a scheduled A-QC analysis could initiate prior to a patient result being released by the operator and result in the pending patient result being overwritten with the A-QC test result.