AMSCO V-PRO 1, AMSCO V-PRO 1 Plus, and AMSCO V-PRO maX Low Temperature Sterilization System (2018-07-23)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer.
Manufacturer
Steris Mexico
790 Avante Ave., Industrial Park
Guadalupe, Nuevo Leon
67190
MEXICO
B. AMSCO V-PRO 1 Plus Low Temperature Sterilization SystemVP20002101
More than 100 numbers, contact manufacturer.
Manufacturer
Steris Mexico
790 Avante Ave., Industrial Park
Guadalupe, Nuevo Leon
67190
MEXICO
C. AMSCO V-PRO maX Low Temperature Sterilization SystemVP30002101
More than 100 numbers, contact manufacturer.
Manufacturer
Steris Mexico
790 Avante Ave., Industrial Park
Guadalupe, Nuevo Leon
67190
MEXICO
A. AMSCO V-PRO 1 Low Temperature Sterilization SystemVP10002101

Introduction

Affected products

A. AMSCO V-PRO 1 Low Temperature Sterilization System
B. AMSCO V-PRO 1 Plus Low Temperature Sterilization System
C. AMSCO V-PRO maX Low Temperature Sterilization System

Reason

The V-PRO 1, V-PRO 1 Plus, and V-PRO maX Low Temperature Sterilization Systems with VAPROX HC Sterilant (DIN# 02303604) are intended for use in the terminal sterilization of properly prepared (cleaned, rinsed and dried) metal and non-metal medical devices in healthcare facilities. The sterilization cycles operate at low pressure and temperature, suitable for processing medical devices sensitive to heat and moisture. Steris has identified that in the unlikely scenario where a punctured cup of VAPROX HC Sterilant expires while the V-PRO Unit (1, 1 Plus, or maX) is powered down, upon power-up the V-PRO Unit will not detect the expired cup of Sterilant; allowing for subsequent cycles to be run. Steris's VAPROX Sterilant microbial efficacy studies support a safety margin for use of the product for three (3) days after the indicated in-use expiry date. Steris has not received any reports of adverse events related to the issue described above.