CVM Material for Calibration Verification of Blood Gas, Electrolyte, Total Hemoglobin, Bilirubin, and Metabolite Systems (2018-07-16) Jul 16, 2018 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Brand name Companies Model or catalog number Lot 350701 and 350702 D. 860/865 BLOOD GAS SYSTEM - CONTROL REAGENTS Manufacturer Siemens Healthcare Diagnostics Inc. 511 Benedict Ave Tarrytown 10591 UNITED STATES 10316535 Lot 350701 and 350702 E. CVM Material Manufacturer Siemens Healthcare Diagnostics Inc. 511 Benedict Ave Tarrytown 10591 UNITED STATES 02147872 Lot 350701 and 350702 B. 840/845 BLOOD GAS SYSTEM - CONTROL REAGENTS Manufacturer Siemens Healthcare Diagnostics Inc. 511 Benedict Ave Tarrytown 10591 UNITED STATES 10316535 Lot 350701 and 350702 C. 348 SYSTEM - REAGENTS Manufacturer Siemens Healthcare Diagnostics Inc. 511 Benedict Ave Tarrytown 10591 UNITED STATES 10316535 Lot 350701 and 350702 A. 850/855 BLOOD GAS SYSTEM - CONTROL REAGENTS Manufacturer Siemens Healthcare Diagnostics Inc. 511 Benedict Ave Tarrytown 10591 UNITED STATES 10316535
Affected products
850/855 BLOOD GAS SYSTEM - CONTROL REAGENTS
840/845 BLOOD GAS SYSTEM - CONTROL REAGENTS
348 SYSTEM - REAGENTS
860/865 BLOOD GAS SYSTEM - CONTROL REAGENTS
CVM Material
Siemens Healthcare Diagnostics has confirmed that the CVM level 1 pCO2 and/or tHb results may recover low within range or lower than expected with the lots 350701 and 350702 of CVM. These lower recoveries may be observed on both the RAPIDPoint and the RAPIDLab Blood Gas Systems. CVM levels 2 through 5 are not affected.