Health Products

IS3000, Redundant Medical Grade Power Supply, Reduced Fan Speed (2018-07-16)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device, Biologic/vaccine
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
RSH0112 (Serial Number)
Manufacturer
Intuitive Surgical
1266 Kifer Road, Building 101,
Sunnyvale
94086
California
UNITED STATES
IS3000, Redundant Medical Grade Power Supply, Reduced Fan SpeedIS3000

Introduction

Affected products

IS3000, Redundant Medical Grade Power Supply, Reduced Fan Speed

Reason

There is an increase in Redundant Medical Grade Power Supply (RMGPS) failures. This is due to a miswiring where the Control/Status and Output wires from the two power supply modules are crossed at the combiner board. If there is an independent failure of one of the power supply modules, the RMGPS shuts down the functioning module, perceiving that it is the failed module due to the crossed wires.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime