IS3000, Redundant Medical Grade Power Supply, Reduced Fan Speed
IS3000
Introduction
Affected products
IS3000, Redundant Medical Grade Power Supply, Reduced Fan Speed
Reason
There is an increase in Redundant Medical Grade Power Supply (RMGPS) failures. This is due to a miswiring where the Control/Status and Output wires from the two power supply modules are crossed at the combiner board. If there is an independent failure of one of the power supply modules, the RMGPS shuts down the functioning module, perceiving that it is the failed module due to the crossed wires.