Health Products

PENTAX Video Duodenoscope (2018-07-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.PENTAX Video Duodenoscope
Manufacturer
Hoya Corporation
6-10-1 NISHI-SHINJUKU
Shinjuku-ku
160-0023
JAPAN
ED34-I10T

Introduction

Affected products

PENTAX Video Duodenoscope

Reason

The complex design of duodenoscopes, including the ED34-i10T, requires meticulous cleaning for effective reprocessing. The US FDA evaluated elevator channel sealing for all duodenoscopes, and required that o-ring compression be increased for all duodenoscopes to be distributed in the US.

PENTAX has made minor design changes in response to this initiative, which are also being made to ED34-i10T units in Canada.