ARCHITECT System - Creatine Kinase Assay (2018-07-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
24358UN18
Manufacturer
Abbott Laboratories Diagnostcis Division
100 Abbott Park Road
Abbott Park
60064
UNITED STATES
ARCHITECT System -  Creatine Kinase Assay7D63-21

Introduction

Affected products

ARCHITECT System - Creatine Kinase Assay

Reason

The purpose of this is to inform customers of a product recall for the ARCHITECT Creatine Kinase (CK) reagent and to provide instructions on what actions must be taken. Abbott has identified a stability issue that may lead to the following issues.
  • Error Code 1054 ""Unable to calculate results, Reaction Check Failure.
  • Quality control (QC) results out of range low.


A potential delay may exist in the reporting of patient results due to quality control values falling outside of expected ranges and the generation of error code 1054. Internal testing on patient samples showed no atypical bias for the impacted reagent lots, therefore if quality controls are in range, patient results are not impacted.