Health Products

ARCHITECT I Sytem - EBV VCA IgM Reagent Kit (2018-07-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
85278LI00
Manufacturer
Abbott GmbH and Co. K.G.
Max Planck Ring 2
Wiesbaden
65205
GERMANY
ARCHITECT I Sytem - EBV VCA IgM Reagent Kit3P66-25

Introduction

Affected products

ARCHITECT I Sytem - EBV VCA IgM Reagent Kit

Reason

ARCHITECT EBV VCA IgM Reagent lots 85278LI00 and 85279LI00 show reduced RLU signals with 4 distributed ARCHITECT EBV VCA IgM Calibrator lots (71146LI00, 73014LI00, 79576LI00 or 83327LI00), leading to decreased cut-off RLU values in assay calibration.

The decreased cut-off RLU values of maximum 20% will result into
  • Quality Control S/CO values shifting up
  • false grayzone patient results in the range of 0.50-0.60 S/CO
  • false reactive patient results in the range 1.00-1.20 S/CO


Cause investigation discovered that the reduced signals are driven by the microparticle component of the reagent, specifically by one lot of microparticle concentrate. Estimated Date of completion: October 2018.

ARCHITECT EBV VCA IgM reagent lots 85278LI00 and 85279LI00 have been placed on Quality Hold on 20-Apr-2018 (QH1019-2018).