X-CAL, Sysmex E-Check, 8Check 3WP X-TRA Control, Sysmex SCS-1000 Universal Calibrator, XN CHECK, XN CAL, XN CAL PF, XN CAL - CUSTOMER SET, XN CAL PF - CUSTOMER SET, and XN-L CHECK (2018-05-08) May 8, 2018 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type III
Lot or serial number Brand name Companies Model or catalog number beginning with 7283 H. XN CAL - CUSTOMER SET Manufacturer Sysmex Corporation 1-5-1 Wakinohama-kaigandori Chuo-ku 651-0073 JAPAN 213527 beginning with 7282 I. XN CAL PF - CUSTOMER SET Manufacturer Sysmex Corporation 1-5-1 Wakinohama-kaigandori Chuo-ku 651-0073 JAPAN 213536 beginning with 7282 F. XN CAL Manufacturer Sysmex Corporation 1-5-1 Wakinohama-kaigandori Chuo-ku 651-0073 JAPAN XN CAL beginning with 7282 G. XN CAL PF Manufacturer Sysmex Corporation 1-5-1 Wakinohama-kaigandori Chuo-ku 651-0073 JAPAN XN CAL PF beginning with 7283 D. Sysmex SCS-1000 Universal Calibrator Manufacturer Sysmex Corporation 1-5-1 Wakinohama-kaigandori Chuo-ku 651-0073 JAPAN 160-4002-0 beginning with 7261 E. XN Check Manufacturer Sysmex Corporation 1-5-1 Wakinohama-kaigandori Chuo-ku 651-0073 JAPAN 213499 beginning with 7269 B. Sysmex E-Check Manufacturer Sysmex Corporation 1-5-1 Wakinohama-kaigandori Chuo-ku 651-0073 JAPAN 199-4004-1 beginning with 7255 C. 8Check 3WP X-TRA Control Manufacturer Sysmex Corporation 1-5-1 Wakinohama-kaigandori Chuo-ku 651-0073 JAPAN 14030040 beginning with 7283 A. X-CAL Manufacturer Sysmex Corporation 1-5-1 Wakinohama-kaigandori Chuo-ku 651-0073 JAPAN 110-4008-0 beginning 7300 J. XN-L CHECK Manufacturer Sysmex Corporation 1-5-1 Wakinohama-kaigandori Chuo-ku 651-0073 JAPAN 213569
Affected products
X-CAL
Sysmex E-Check
8Check 3WP X-TRA Control
Sysmex SCS-1000 Universal Calibrator
XN Check
XN CAL
XN CAL PF
XN CAL - CUSTOMER SET
XN CAL PF - CUSTOMER SET
XN-L CHECK
Sysmex America, Inc. received and increased number of complaints for low recovery or imprecision of Platelet (PLT) counts and associated parameters including Mean Platelet Volume (MPV), and Immature Platelet Fraction (IPF) with hematology controls and calibrators. Remixing the product and repeating analysis gave acceptable results. No product defect was identified; however, mixing the product per the package insert instructions was insufficient to ensure consistent PLT recovery within the assay sheet ranges.