T2 Proximal Humeral Nail System - Long Nail (2018-06-15)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
K0ACFED
Manufacturer
Stryker Trauma GMBH
Prof.-Kuntscher-Str.1-5
Schoenkirchen
24232
GERMANY
T2 Proximal Humeral Nail System - Long Nail1832-3826S

Introduction

Affected products

T2 Proximal Humeral Nail System - Long Nail

Reason

The manufacturer has determined that products have left the factory where the quality of the sealed seam of the sterile packaging did not correspond to the specification. During the sealing process, a tyvek lid stuck to the sealing plate. The subsequently packaged products were sealed with the lid sticking to the seal plate.
Though testing of the sealed seam confirmed that the integrity of the sterile barrier is intact, integrity of the sterile barrier cannot be assured over 5 years shelf life.