ARCHITECT c8000, c16000, c4000 Processing Module (2018-07-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Abbott Laboratories Diagnostcis Division
100 Abbott Park Road
Abbott Park
60064
UNITED STATES
B. ARCHITECT c16000 Processing Module03L77-01
More than 10 numbers, contact manufacturer.
Manufacturer
Abbott Laboratories Diagnostcis Division
100 Abbott Park Road
Abbott Park
60064
UNITED STATES
C. ARCHITECT c4000 Processing Module02P24-01
More than 10 numbers, contact manufacturer.
Manufacturer
Abbott Laboratories Diagnostcis Division
100 Abbott Park Road
Abbott Park
60064
UNITED STATES
ARCHITECT c8000 Processing Module01G06-01

Introduction

Affected products
  1. ARCHITECT c8000 Processing Module
  2. ARCHITECT c16000 Processing Module
  3. ARCHITECT c4000 Processing Module

Reason

If architect pressure monitoring is not communicating properly during power up, the system will notify the user with a pop-up message including error codes 3580 to 3585 or 5381. The architect operations manual section 10 provides instructions on how to correctly resolve these errors. If the user chooses to transition the architect to the running status without successfully resolving error codes 3580 to 3585 or 5381, there is the potential to generate incorrect results as the pressure monitoring system is inactive.