Health Products

Carestation 650 (2018-07-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Datex-Ohmeda, Inc.
3030 Ohmeda Drive, P.O. Box 7550
Madison
53707-7550
Wisconsin
UNITED STATES
Carestation 6501012-9650-000

Introduction

Affected products

Carestation 650

Reason

On systems equipped with the Auxiliary Common GGAS Utlet (ACGO), there is a potential for the auxiliary gas switch to unexpectedly switch from the ACGO position back to the circle position. On systems equipped with the Aux O2 + Air function the switch could unexpectedly switch from the Aux O2 + Air position back to the circle position. These situations could lead to hypoxia. It would still be possible to ventilate the patient using the circle breathing system in the manual bag or mechanical ventilation modes.