Health Products

BIS Complete Monitoring System (2018-07-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer.
Manufacturer
Covidien LLC
15 Hampshire Street
Mansfield
02048
Massachusetts
UNITED STATES
BIS Complete Monitoring System186-1046

Introduction

Affected products

BIS Complete Monitoring System

Reason

Replacement instructions were revised for the Li-Ion battery packs used in BIS Vista Monitoring System. These revised instructions will require battery pack replacement after 4 years due to the potential for aged battery packs to experience an internal short that may result in a thermal event.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime