Health Products

ADVIA Centaur and ADVIA Centaur CP System - Creatine Kinase-MB (CKMB) Assay (2018-07-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
57834A64
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
B. ADVIA Centaur CP System-Creatine Kinase-MB (CKMB) Assay9318028
57834A64
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
A. ADVIA Centaur System-Creatine Kinase-MB (CKMB) Assay9318028

Introduction

Affected products
  1. ADVIA Centaur System-Creatine Kinase-MB (CKMB) Assay
  2. ADVIA Centaur CP System-Creatine Kinase-MB (CKMB) Assay

Reason

Siemens Healthcare Diagnostics has identified a bias with ADVIA Centaur CKMB Calibrator Kit Lots ending in 64 (ck64) when compared to internal standards and previously released ADVIA Centaur CKMB Calibrator Kit Lots ending in 63 (ck63) on the ADVIA Centaur/XP/XPT and ADVIA Centaur CP systems.
Alignment to the internal standardization is restored with the release of ADVIA Centaur CKMB Calibrator Kit Lots ending in 68 (ck68). Customers may observe a shift in Quality Control (QC) material, Master Curve Material (MCM) and patient results when transitioning from ADVIA Centaur CKMB Calibrator Kit Lots ending in 64. Siemens has also identified the potential for out of range ADVIA Centaur CKMB Master Curve Material results when using the lot 35061on the ADVIA Centaur/XP/XPT and ADVIA Centaur CP systems.