Health Products

da Vinci S/Si and da Vinci X/Xi EndoWrist 8mm Monopolar Curved Scissors (2018-08-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
N10160316B. da Vinci X/Xi EndoWrist 8mm Monopolar Curved Scissors
Manufacturer
Intuitive Surgical Inc.
1266 Kifer Road, Building 100, 101, 102, 103
Sunnyvale
94086
California
UNITED STATES
470179
N10170819A. da Vinci S/Si EndoWrist 8mm Monopolar Curved Scissors
Manufacturer
Intuitive Surgical Inc.
1266 Kifer Road, Building 100, 101, 102, 103
Sunnyvale
94086
California
UNITED STATES
420179

Introduction

Affected products
  1. da Vinci S/Si EndoWrist 8mm Monopolar Curved Scissors
  2. da Vinci X/Xi EndoWrist 8mm Monopolar Curved Scissors

Reason

Specific versions of the da Vinci S/Si and da Vinci X/Xi Monopolar Curved Scissor (MCS) instruments that may be more susceptible to developing cracks near the proximal edge of the extension tube, potentially attributed to mishandling and misuse as defined by the Instrument and Accessories User Manual. These cracks may fracture the plastic surrounding the instrument shaft, and may cause fragments of the extension tube to fall, including potentially into the patient.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime