Health Products

UniCel DxH 600, 800, and 900 Coulter Cellular Analysis System (2018-08-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Software 1.3.0 and below
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
B. UniCel DxH 600 Coulter Cellular Analysis SystemB23858
All lots.
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
C. UniCel DxH 900 Coulter Cellular Analysis SystemC11478
Software 3.2.0 and below
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
A. UniCel DxH 800 Coulter Cellular Analysis System629029

Introduction

Affected Products
  1. UniCel DxH 800 Coulter Cellular Analysis System
  2. UniCel DxH 600 Coulter Cellular Analysis System
  3. UniCel DxH 900 Coulter Cellular Analysis System

Reason

Beckman Coulter has confirmed complaints of sporadic erroneously elevated platelet results without flags or system messages. Other parameters are not affected by this issue. The platelet parameter may be erroneously elevated without flags or system messages. Erroneously elevated results could be reported out of the laboratory.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime