Health Products

Formula Resector Cutter, Formula Aggressive Plus Cutter and Formula Barrel Bur (2018-08-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
8069955B. Formula Aggressive Plus Cutter
Manufacturer
Stryker Sustainability Solutions, Inc.
1810 West Drake Dr.
Tempe
85283
Arizona
UNITED STATES
375-544-000
8121786C. Formula Barrel Bur
Manufacturer
Stryker Sustainability Solutions, Inc.
1810 West Drake Dr.
Tempe
85283
Arizona
UNITED STATES
375-941-012
8082746A. Formula Resector Cutter
Manufacturer
Stryker Sustainability Solutions, Inc.
1810 West Drake Dr.
Tempe
85283
Arizona
UNITED STATES
375-542-000

Introduction

Affected products
  1. Formula Resector Cutter
  2. Formula Aggressive Plus Cutter
  3. Formula Barrel Bur

Reason

Stryker's sustainability solutions division has determined that affected lots of reprocessed arthroscopic shavers may display the incorrect catalog number and speed settings when connected to the generator, require manual mode for use on core generators, or display an error message on crossfire 1 and 2 generators which does not allow the device to be used.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime