Health Products

LIAISON Control Measles IgM (2018-08-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
175012
Manufacturer
Diasorin Inc.
1951 Northwestern Avenue, P.O. Box 285
Stillwater
55082
Minnesota
UNITED STATES
LIAISON Control Measles IgM318821

Introduction

Affected products

LIAISON Control Measles IgM

Reason

Positive controls, Part nos. 3500090 and 3500100, in the affected kit lots may result high out of the certificates of analysis range when tested on LIAISON xl in conjunction with liaison measles IgM kit (ref 318820) lot nos. 174017 and 174018. If this combination of kit and control produces a valid run (i.e. calibrators and controls are within established ranges), results can be reliably reported.