Health Products

Citadel Plus (2018-08-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
P0423627
Manufacturer
ArjoHuntleigh AB
Hans Michelsensgatan 10
Malmo
21120
SWEDEN
Citadel PlusFXXXXXXXALVBB

Introduction

Affected products

Citadel Plus

Reason

It has been identified that the power drive function of the citadel plus bed may potentially be operated with only the left-handle safety switch depressed on the condition that the bed is driven forward and encounters heavy resistance (i.e. steep slopes in excess of greater than 7 degrees). This is due to a software deficiency that has been detected after receiving several customer complaints.