Health Products

HEMA-TEK Slide Stainer - Stain Pak - Modified Wright's Stain and Wright-Giemsa Stain (2018-08-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
439684
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
B. HEMA-TEK Slide Stainer - Stain Pak - Wright-Giemsa Stain01178588
433521
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New York
UNITED STATES
A. HEMA-TEK Slide Stainer - Stain Pak - Modified Wright's Stain06689653

Introduction

Affected products

A. HEMA-TEK Slide Stainer - Stain Pak - Modified Wright's Stain
B. HEMA-TEK Slide Stainer - Stain Pak - Wright-Giemsa Stain

Reason

Manufacturer communication indicating that the lots of stain included in the notification may form precipitate on a slide during the staining process. This precipitation may make it more difficult to successfully perform peripheral blood microscopy procedures. This problem has been linked to a single lot of raw material used in manufacturing.