Health Products

Philips SureSigns VM4, VM6, VM8, VS3, VSV (Vital Signs Viewer) Central Station, VS2 Vital Signs Monitor, and VS4 Vital Signs Monitor System - Vital Signs Monitor (2018-07-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 100 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
UNITED STATES
F. SureSigns VS2 Vital Signs Monitor863080
More than 100 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
UNITED STATES
G. SureSigns VS4 Vital Signs Monitor System - Vital Signs Monitor863283
More than 100 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
UNITED STATES
D. SureSigns VS3863071
More than 100 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
UNITED STATES
E. SureSigns VSV (Vital Signs Viewer) Central Station863067
More than 100 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
UNITED STATES
B. SureSigns VM6863064
More than 100 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
UNITED STATES
C. SureSigns VM8863066
More than 100 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
UNITED STATES
A. SureSigns VM4863063

Introduction

Affected products

A. SureSigns VM4
B. SureSigns VM6
C. SureSigns VM8
D. SureSigns VS3
E. SureSigns VSV (Vital Signs Viewer) Central Station
F. SureSigns VS2 Vital Signs Monitor
G. SureSigns VS4 Vital Signs Monitor System - Vital Signs Monitor

Reason

Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited.
These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles.
Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime