G. SureSigns VS4 Vital Signs Monitor System - Vital Signs Monitor
863283
More than 100 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
UNITED STATES
D. SureSigns VS3
863071
More than 100 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
UNITED STATES
E. SureSigns VSV (Vital Signs Viewer) Central Station
863067
More than 100 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
UNITED STATES
B. SureSigns VM6
863064
More than 100 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
UNITED STATES
C. SureSigns VM8
863066
More than 100 numbers, contact manufacturer.
Manufacturer
Philips Medical Systems
3000 Minuteman Road
Andover
01810
UNITED STATES
A. SureSigns VM4
863063
Introduction
Affected products
A. SureSigns VM4
B. SureSigns VM6
C. SureSigns VM8
D. SureSigns VS3
E. SureSigns VSV (Vital Signs Viewer) Central Station
F. SureSigns VS2 Vital Signs Monitor
G. SureSigns VS4 Vital Signs Monitor System - Vital Signs Monitor
Reason
Philips has received several reports in which monitors with lithium ion batteries that had exceeded their battery life expectancy have overheated or ignited.
These batteries should be replaced every 3 years or upon reaching 300 charge-discharge cycles.
Although the Philips SureSigns VS & VM Monitors and View Station (VSV) can display actual information on either or both battery age and charge-discharge cycles, the existing labeling does not include full instructions on how to use this information to determine when to replace the battery.