VITROS 5600 Integrated System (2018-08-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
VITROS 5600 Integrated System6802413

Introduction

Affected products

VITROS 5600 Integrated System

Reason

Under very specific circumstances, incorrect positioning of the sample metering proboscis on a VITROS 5600 Integrated System may cause an insufficient sample volume to be dispensed onto a microslide without suppressing the potentially incorrect result. If these specific circumstances occur, the proboscis may deposit fluid on the tip locator or top surface of the slide and transfer sample fluid into the incubator. Mis-positioning of the proboscis may lead to an insufficient volume of sample being dispensed onto a microslide and may lead to incorrect results.