Health Products

Oxylog 3000 (2018-07-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
DRAEGERWERK AG & CO
MOISLINGER ALLEE 53-55
LUBECK
23542
GERMANY
Oxylog 30002M86955

Introduction

Affected products

Oxylog 3000

Reason

A new software is now available with immediate effect for the Oxylog 3000. The new SW version 1.24 resolves a programming defect in the currently available SW version 1.23; The error resulted in the following three user advisory messages no longer being reported:
  • 21 % by vol. Oxygen
  • Self-test O.K.
  • Setting not confirmed


The new SW version 1.24 ensures that all user advisory messages are correctly displayed again. Other Oxylog ventilators, including the Oxylog 3000 plus, do not show this defective response.