Health Products

Siemans Dimension Software Version 10.2.2 (2018-06-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
H. Dimension Xpand System - Xpand Plus With HM Analyzer - Class 2765000.931
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
I. Imension Xpand System - Xpand Plus With HM Analyzer765000.931
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
F. Dimension Xpand Plus Basic Analyzer - Class 2765000.921
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
G. Dimension EXL System - Dimension EXL 200 Analyzer778041.911
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
D. Dimension Chemistry System - RxL Max Analyzer Class 3752000.901
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
E. Dimension EXL System - Dimension EXL Analyzer With LM759020.911
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
B. Dimension EXL System - Dimension EXL Analyzer With Loci Module759020.911
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
C. Dimension Clinical Chemistry System - RxL Max Analyzer - Class 2752000.901
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
A. Dimension Xpand System - Xpand Plus Basic Analyzer765000.921
More than 10 numbers, contact manufacturer.
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
J. Dimension Chemistry System - Dimension RxL Max W/HM Analyzer752000.911

Introduction

Affected products
  1. Dimension Xpand System - Xpand Plus Basic Analyzer
  2. Dimension EXL System - Dimension EXL Analyzer With Loci Module
  3. Dimension Clinical Chemistry System - RxL Max Analyzer - Class 2
  4. Dimension Chemistry System - RxL Max Analyzer Class 3
  5. Dimension EXL System - Dimension EXL Analyzer With LM
  6. Dimension Xpand Plus Basic Analyzer - Class 2
  7. Dimension EXL System - Dimension EXL 200 Analyzer
  8. Dimension Xpand System - Xpand Plus With HM Analyzer - Class 2
  9. Imension Xpand System - Xpand Plus With HM Analyzer
  10. Dimension Chemistry System - Dimension RxL Max W/HM Analyzer

Reason

Siemens is informing users of the potential for intermittent failures with the group calibration feature in Dimension Software Version 10.2.2. A calibration may not auto accept when performing a group calibration with more than one Quality Control (QC) product. If all QC levels are not processed, Dimension QCC PowerPak auto acceptance for that calibration will not complete.