Health Products

Paradigm Insulin Infusion Pumps Remote Programmer (2018-08-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
MEDTRONIC MINIMED
18000 DEVONSHIRE ST
NORTHRIDGE
91325
California
UNITED STATES
Paradigm Insulin Infusion Pumps Remote ProgrammerMMT-503

Introduction

Affected products

Paradigm Insulin Infusion Pumps Remote Programmer

Reason

An external security researcher has identified a potential vulnerability related to the MiniMed Paradigm Family of insulin pumps and corresponding remote controller. The researcher's report states that an unauthorized individual in close proximity of an insulin pump user could potentially copy the wireless Radio Frequency (RF) signals from the user's remote controller (while they are in the process of delivering a remote bolus) and play those back later to deliver an involuntary bolus of insulin to the user. This could lead to potential health risks such as hypoglycemia if additional insulin is delivered beyond the user's insulin requirements.