Health Products

Leica M525 F20 Microscope Systems (2018-08-17)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.Leica M525 F20 Microscope Systems
Manufacturer
Leica Microsystems (Schweiz) Ag
Max Schmidheiny-Strasse 201
Heerbrugg
9435
SWITZERLAND
10448284

Introduction

Affected products

Leica M525 F20 Microscope Systems

Reason

Leica Microsystems received five customer complaints stating that users inadvertently tipped over their Leica M525 F20 Microscope System. In four cases, it was confirmed, that the issue occurred when users handled the microscope while the swing arm was in fully extended position. The five complaints occurred in five different countries over a time frame between the years 2010 and 2017.