VITROS ECI Immunodiagnostic Products Vitamin B12 Reagent Pack Controls & CA (2018-08-31)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
2601
Manufacturer
Ortho Clinical Diagnostics
Felindre Meadows
Pencoed, Bridgend
CF35 5PZ
UNITED KINGDOM
VITROS ECI Immunodiagnostic Products Vitamin B12 Reagent Pack Controls & CA145 3489

Introduction

Affected products

VITROS ECI Immunodiagnostic Products Vitamin B12 Reagent Pack Controls & CA

Reason

Ortho Clinical Diagnostics confirmed that VITROS Vitamin B12 Reagent Pack 1/2, Lot 2601 exhibits the potential for an increased frequency of calibration failures. When a calibration failure occurs, customers have reported that their VITROS System generated the following condition codes to indicate that the calibrator signal index exceeds the limit:
  • For VITROS 3600/5600 Systems: PW7-012 (Calibrator signal index above limit Level %d)
  • For VITROS ECi/ECiQ Systems: 542-037 (542-037: Calibration %s Lot %s failed)


Our preliminary testing indicates that the calibration failure is related to the storage of a component used during the manufacture of some VITROS Vitamin B12 Reagent Packs 1/2, Lot 2601 and is not associated with VITROS Vitamin B12 Calibrators. Note: Not all reagent packs from Lot 2601 are affected.