DSX- Revelation software 6.08-6.28 (2018-08-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Dynex Technologies Inc.
14340 Sullyfield Cir
Chantilly
20151-1621
Virginia
UNITED STATES
DSX- Revelation software 6.08-6.2865078-XXX

Introduction

Affected products

DSX- Revelation software 6.08-6.28

Reason

The DSX analyzer utilizes operating software Revelatlon. This software use set up as default with an algorithm for dealing with over/ under conversion. If an end user turns this off and they use the over limit in their Quality Control Equation the software will place 0.00 in the equation leading to an incorrect QC interpretation and potential incorrect value assigned to the result.