Health Products

Mitek Screw and Washer Depth Gauge (2018-08-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Depuy Mitek
325 Paramount Drive
Raynham
02767
Massachusetts
UNITED STATES
Mitek Screw and Washer Depth Gauge219024

Introduction

Affected products

Mitek Screw and Washer Depth Gauge

Reason

DePuy Mitek has received reports of the Mitek Screw and washer depth gauge measuring incorrectly. An investigation determined the depth gauge has an offset in the design that result in a reading approximately 3.2mm shorter than the actual length of the bone hole. This field safety notice is for the removal of all lots.