Activa PC+S Implantable Neurostimulator (INS) (2018-08-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
NKM734554HActiva PC+S Implantable Neurostimulator (INS)
Manufacturer
Medtronic Inc.
710 Medtronic Parkway
Minneapolis
55432
Minnesota
UNITED STATES
37604

Introduction

Affected products

Activa PC+S Implantable Neurostimulator (INS)

Reason

Manufacturer identified anomaly in the Activa PC+S firmware. This anomaly can result in a Model 37604 Activa PC+S Implantable Neurostimulator (INS) that cannot communicate with programming instruments, such as the Model 37642 patient programmer or Model 8840 clinician programmer. The issue also affects Activa PC+S devices that have been updated with the investigational nexus-e firmware.