HeartWare Ventricular Assist System - HeartWare Controller (2018-08-29)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer.
Manufacturer
Heartware Inc.
14400 NW 60th Avenue
Miami Lakes
33014
Florida
UNITED STATES
HeartWare Ventricular Assist System - HeartWare Controller1407CA

Introduction

Affected products

HeartWare Ventricular Assist System - HeartWare Controller

Reason

Controller units may not meet the labeled standard for protection against water or fluid ingress. Specifically, some units have the potential to develop hairline cracks at the power ports of the controller housing. Hairline cracks may allow for water ingress if the patient does not follow the instructions for use and the patient manual, including use of a water-resistant shower bag to help protect the unit.
If a controller develops a hairline crack and is unprotected from water or fluid exposure as described in the IFU, the unit may experience water or fluid ingress. This can lead to varying degrees of controller malfunction, including pump stop.