Health Products

Giraffe, Panda Freestanding, and Panda iRes Warmer (2018-09-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.B. Panda Freestanding Warmer
Manufacturer
Datex-Ohmeda, Inc
9900 Innovation Drive
Wauwatosa
53226
Wisconsin
UNITED STATES
2063568-001
More than 1000 numbers, contact manufacturer.C. Panda iRes Warmer
Manufacturer
Datex-Ohmeda, Inc
9900 Innovation Drive
Wauwatosa
53226
Wisconsin
UNITED STATES
M1112198
More than 100 numbers, contact manufacturer.A. Giraffe Warmer
Manufacturer
Datex-Ohmeda, Inc
9900 Innovation Drive
Wauwatosa
53226
Wisconsin
UNITED STATES
M1118179

Introduction

Affected products
  1. Giraffe Warmer
  2. Panda Freestanding Warmer
  3. Panda iRes Warmer

Reason

GE Healthcare has found an opportunity to improve performance of its giraffe and panda warmers. This is achieved by switching to a different type of fuse on the power board. Warmers manufactured prior to September 30, 2017 utilized an end cap made of silver. Over time, corrosion/oxidation of the silver end cap may result to non-typical (high) contact resistance between the cap surface and the fuse holder. The new fuses have an end cap made of gold. The gold cap will mitigate corrosion.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime