Health Products

RADREX-i (2018-09-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
W8B14X2014
Manufacturer
Toshiba Medical Systems Corporation
1385 Shimoishigami
Otawara-Shi, Tochigi-ken
324-8550
JAPAN
RADREX-iDRAD-3000E/W8

Introduction

Affected products

RADREX-i

Reason

When the user performed radiography with wireless X-Ray Detector TFP-4336W, a message window was displayed on a monitor that said image transmission had not completed with no radiographic image. It also showed "ok" button to reacquire image data from the x-ray detector and "cancel" button to cancel the reacquisition. The user selected "ok" button, but the same message window appeared again. The user repeated the same operations several times and finally he/she selected "cancel" button to quit the re-acquisition mode. After system was restarted, the system worked normally.