Health Products

CentriMag Motor (2018-09-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
Not applicableCentriMag Motor
Manufacturer
Thoratec Switzerland GmbH
Technoparkstrasse 1
Zürich
8005
SWITZERLAND
201-10002

Introduction

Affected products

CentriMag Motor

Reason

Abbott has received reports where damage to the cable that connects the motor to the console resulted in interruption of active support. Analysis of devices associated with the motor stop field complaints showed that the motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection.