NEURO OMEGA
Introduction
Affected products
Neuro Omega system
Reason
In case Neuro Omega "Global stimulation return" connector of the EMG/ECOG Headbox is not properly connected, stimulation current may return through the Microelectrodes' contacts (Micro and Macro). Current can reach high charge density which may cause tissue harm.