BonD-MaX and BOND-III System (2018-09-13)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Leica Biosystems Melbourne Pty. Ltd.
495 Blackburn Road
Mount Waverley, Victoria
3149
AUSTRALIA
B. BOND-III SystemNot applicable
Not applicable
Manufacturer
Leica Biosystems Melbourne Pty. Ltd.
495 Blackburn Road
Mount Waverley, Victoria
3149
AUSTRALIA
A. BonD‑MaX SystemNot applicable

Introduction

Affected products

A. BonD‑MaX System
B. BOND-III System

Reason

This Field Action Notice applies to the Zebra Power Supply Unit (PSU) for the Zebra printer models Zebra TLP3842 (affected PSU part number 808113-001 or P1028888-001) and Zebra GX430t (affected PSU part number 808101-001. These were supplied by Leica Biosystems as part of the bond slide labeler for use with bond-iii and bond-max instruments.

The affected models of Zebra printers, which serve as the external power source for these printers, have been identified as potentially containing an improperly mixed phosphorus compound that is used in the flame retardant resin in the DC power supply connectors. This deficiency in the phosphorus based connector resin, when combined with moisture and humidity over time, can potentially lead to overheating or a fire hazard.