SYNCHRON Systems Lipase (LIP) Reagent (2018-09-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
M712448
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
SYNCHRON Systems Lipase  (LIP) Reagent476851

Introduction

Affected products

SYNCHRON Systems Lipase  (LIP) Reagent

Reason

Control and patient samples tested with lip reagent lot M712448 may recover lower than those tested with other lots. Samples below 100 U/L may recover up to 10.7 U/L lower. Samples above 100 U/L may recover up to 8.9% lower. Lower recovery of lip in control samples may exceed the lower limits of a laboratory's established QC range or published insert range.