Alinity i Processing Module (2018-09-18)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Ai01001 to Ai01336
Manufacturer
Abbott GmbH and Co. K.G.
Max Planck Ring 2
Wiesbaden
65205
GERMANY
Alinity i Processing Module03R65-01

Introduction

Affected products

Alinity i Processing Module

Reason

Abbott has identified an issue with the Alinity i bulk solution Dispense Pumps (part number A-35001280-01) resulting in pump failures. The issue is linked to the design of an internal seal within the pump that allows a small amount of leakage of bulk solution(s) that eventually results in the pump failure. The amount of the leakage does not impact the ability of the pump to accurately dispense bulk solutions, and the leakage is contained within the service-only area of the instrument. Therefore, this issue does not pose a risk of incorrect results or a risk of physical exposure to the operator. The failure of this pump could result in unexpected instrument down time potentially causing a delay in generating results.