Health Products

BD FocalPoint GS Review Station, Slide Profiler, and FocalPoint Slide Profiler (Refurbished) (2018-09-20)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Becton Dickinson and Company
1 Becton Drive
Franklin Lakes
07417
UNITED STATES
B. FocalPoint Slide Profiler490652
FP00178R
Manufacturer
Becton Dickinson and Company
1 Becton Drive
Franklin Lakes
07417
UNITED STATES
C. FocalPoint Slide Profiler (Refurbished)490398
144520
Manufacturer
Becton Dickinson and Company
1 Becton Drive
Franklin Lakes
07417
UNITED STATES
A. BD FocalPoint GS Review Station490189

Introduction

Affected products
  1. BD FocalPoint GS Review Station
  2. FocalPoint Slide Profiler
  3. FocalPoint Slide Profiler (Refurbished)

Reason

The BD FocalPoint GS Review Station and BD FocalPoint Slide Profiler may not function at the lower range (<15°C) of the labeled operating temperatures. In addition, it was determined that testing facilities do not guarantee the ability to maintain %RH below 20%. The instrument labeling has been revised to indicate an operating temperature lower limit of 18°C and operating humidity of 20 to 80% RH.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime