Zilver 518 and 635 Biliary Stent (2018-09-26)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer.Zilver 518 and 635 Biliary Stent
Manufacturer
COOK IRELAND LTD.
O'Halloran Road, National Technological Park
Limerick
IRELAND
ZIB5-125-4.0-20

Introduction

Affected products

Zilver 518 and 635 Biliary Stent

Reason

This recall is being initiated for specific lots of the the the Zilver 518 Biliary Stent and Zilver 635 Biliary Stent devices. We are initiating a voluntary correction to the instructions for use (IFU) for these stent devices, because the instructions for removing the delivery system following stent deployment are incorrect.