Beckman Coulter DxC 700 AU Clinical Chemistry Analyzer (2018-06-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Software version 1.07
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
UNITED STATES
Beckman Coulter DxC 700 AU Clinical Chemistry AnalyzerB86443

Introduction

Affected products

Beckman Coulter DxC 700 AU Clinical Chemistry Analyzer

Reason

A software limitation occurs with software version 1.07 when using the analyzer maintenance help, the ise maintenance help or the consumables help icons. The issue may cause an incorrectly displayed dialog box or white screen. In some cases the system has to be rebooted. This could lead to a delay of reporting patient results.