Dimension EXL Analyzer with LM, Dimension EXL 200 Analyzer, and Dimension RxL Max w/HM Analyzer (2018-10-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
DR271567B. Dimension EXL 200 Analyzer
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
778041.911
220686C. Dimension RxL Max w/HM Analyzer
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
752000.911
12252027Dimension EXL Analyzer with LM
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES
759020.911

Introduction

Affected products
  1. Dimension EXL Analyzer with LM
  2. Dimension EXL 200 Analyzer
  3. Dimension RxL Max w/HM Analyzer

Reason

Siemens has identified that some thermal chamber motors have electrical cables with single insulation rather than the required double insulation. There is no exposed current due to the single insulation. The thermal chamber motor is located under the thermal chamber within the dimension instrument and can be accessible when changing the thermal chamber filter.